The R15 rack accommodates 15 × 10 mL or 15 × 20 mL vials for various sample types and techniques. The R60 is an expanded rack that accommodates up to 60 vials.
For 1.5/2 mL headspace vials, the VT54 is available and accommodates up to 54 vials.
When working with crimp-top vials fitted with HiSorb septa, HiSorb septum plugs (C-HSPVLP-20) are available to seal the vial after probe extraction. These are recommended for samples that contain odorous or hazardous compounds and/or high levels of VOCs that could contaminate lab air after sampling is complete, and can be fitted manually or using automation. To enable automation, the RVLP rack is available and can accommodate up 20 plugs.
Centri has a number of user-installable consumable parts that should be checked and replaced regularly as part of a routine maintenance plan. These include injector liners, liner seal O-rings and focusing traps.
When Centri is fitted with a HiSorb wash station, the sparge filter must also be checked. When fitted with a standard wash module, the vial/snap-cap and seal will need to be checked. When fitted with a tube module, DiffLok caps, tubes and trays will need to be checked.
All additional parts (not mentioned on this page) should be checked and replaced by Markes-trained/approved engineers as part of a maintenance agreement.
Spare injector caps and HiSorb handles are available in case of accidental loss or damage, or when undergoing a system upgrade/expansion.